From ResMed site:Clinical Use
Disassemble the mask according to the instructions in the User Guide.
Cleaning
1. Clean the individual parts of the mask with a soft bristle brush for one minute while soaking
them in detergent or disinfectant1. Pay particular attention to all crevices and cavities.
2. Rinse the components twice by agitating them vigorously in drinking quality water (5 litres per
mask).
1 Cleaning: ResMed has tested the disinfectant Alconox (diluted at 1%).
1. Soak the disinfectable mask components in a commercially available solution of orthophthalaldehyde
0.55% (eg, CIDEX™ OPA) or glutaraldehyde 3.4% (eg, CIDEX Plus™)-both Johnson&Johnson products.
2. Rinse the mask components in drinking quality water (5 litres per mask) and allow them to air
dry out of direct sunlight.
2 Disinfection: ResMed masks have been validated, for the given number of cycles, using solutions of ortho-phthalaldehyde (0.55%
CIDEX OPA for 20 minutes) or glutaraldehyde (3.4% CIDEX Plus for 20 minutes).
Handwash the headgear in warm 86°F (30°C) water using mild soap. Rinse well and allow it to air dry out of
direct sunlight.